Thermo Scientific 5030 Instruction Manual page 37

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Thermo Fisher Scientific
To continue with the acceptance testing, scroll forward to the Service menu
and press YES to enter this menu. Once there, press SET to unlock the
keypad and you should see the following screen:
T1: Ambient Temperature (deg C),
T2: Sample Temperature (deg C),
T3: Orifice Temperature (deg C),
T4: Heater External Wall Temperature (deg C),
P1: Orifice Pressure Drop (hPa),
P2: Vacuum under sample filter tape (hPa),
P3: Barometric Pressure (hPa),
RH: Sample Relative Humidity (% RH).
This is the Service menu. Press NEXT.
This is the Calibration menu. Press NEXT.
You have now returned to the main User Screen.
By pressing NEXT or BACK you may scroll to the
Operation, Service, and Calibration menus. By
pressing DISP, you will scroll through the User
Screen sub-menus again.
From here, continue to press NEXT until you see
the following screen:
Press YES to enter this Service sub-menu and
continue to press NEXT until you see the
following screen:
At this screen, the top row is an abbreviated text of
the digital measurements being provided. The
description of these abbreviations is as follows:
Model 5030 SHARP Monitor Instruction Manual