Air Pressure is Low
The air pressure is displayed on the front of the services module. If the air pressure
is correct then the display will be green. If the air pressure is too low or too high
then the display will be red. It's likely that if the display is red then the compressor
will also be running.
If the compressor runs for more than 5 minutes and air pressure hasn't been
restored to normal, check the air line between the services module and the Sorter
for leaks.
If the air line appears OK, or the compressor is not running and the pressure is still
not restored, contact Abbott Customer Service.
Auxiliary air can be connected to the services module from the building or a bottle
supply via the auxiliary air input. This will allow the PathFinder 350S to continue
running while waiting for the services module to be serviced. Contact your facilities
manager to organize an alternate supply.
8.6
Hints and Tips
This section contains any operational hints and tips that may optimise the
performance and reliability of the PathFinder 350S.
Page 104
Figure 8-11 BIOS Advanced Chipset Menu - Workcell
In the list that is displayed change the display device setting to CRT, (NOT
LCD).
Press F10 to save the changes and restart.