Power off the services module, wait a few seconds, and power back on. If still
nothing is displayed then contact Abbott Customer Service.
8.5.7
The LCD Monitor Loses Signal After Windows Starts
If the monitor displays No Signal soon after Windows begins to load, the workcell
controller BIOS settings have been changed. This is easily fixed but requires a
keyboard to be attached to the back of the Sorter:
1)
2)
3)
4)
03P27-42 Rev A
Switch off the PathFinder 350S and attach a USB keyboard to the back panel
of the Sorter.
Switch on the services module and watch the workcell controller power on.
Immediately start pressing the Delete key. Repeat pressing Delete every
second or so until the BIOS screen is displayed. It will appear as shown in
Figure 8-10.
Figure 8-10 BIOS Menu - Workcell Controller PC
Navigate the BIOS menu using the keyboard arrow keys.
Select the Advanced Chipset Features menu.
© Copyright 2010 Aim Lab Pty Ltd