Aisys

3.12 Power failure test

3.13 Electrical safety tests

3-12
1. Connect the power cord to a wall outlet. The mains indicator on the front
panel comes on when AC Power is connected.
2. Set the system switch to On and Start a case.
3. Unplug the power cord with the system turned on.
4. Make sure that the power failure alarm comes on.
5. Make sure the following message is displayed:
• Plug in power cable. On battery
6. Connect the power cable again.
7. Make sure the alarm cancels.
Make sure the system is completely assembled and all accessory devices are
connected to electrical outlets.
1. Connect an approved test device (e.g. UL, CSA, or AAMI) and verify that the
leakage current is less than:
Voltage
120/100 Vac
220/240 Vac
2. Make sure that the resistance to ground is less than 0.2Ω between an
exposed metal surface and the ground pin on the power cord.
Max. Leakage Current
300 µAmps
500 µAmps
04/08 M1046983