GE Datex-Ohmeda Aisys Carestation Reference Manual page 161

Anesthesia machine
Table of Contents

Advertisement

M1046983 04/08
4. Verify the "Schedule Service Calibration" message is not present in the
normal display.
5. Complete the System Checkout by performing the following steps:
a. Inspect the system (TRM - Section 3.1)
b. System checkout (TRM - Section 3.2)
Note: You must insert a Test Cassette for the Machine Check - System
Agent Delivery Test to run the extended diagnostic Vaporizer Test.
c. Pipeline and Cylinder tests (TRM - Section 3.6)
d. Flush Flow test (TRM - Section 3.7)
e. Alarm tests (TRM - Section 3.8)
f. Alternate O
flowmeter tests (TRM - Section 3.9)
2
g. Auxiliary O
flowmeter tests, if equipped with option (TRM - Section
2
3.10)
h. Integrated suction regulator tests, if equipped with option
(TRM - Section 3.11)
i. Power failure test (TRM - Section 3.12)
j. Electrical safety tests (TRM - Section 3.13)
6 Installation and Maintenance
6-3