GE Datex-Ohmeda Aisys Carestation Reference Manual page 216

Anesthesia machine
Table of Contents

Advertisement

Aisys
Alternate O
/ System Malfunction Screen Troubleshooting
2
Start
Download and combine the
Error / Alarm Logs.
(Section 12.6.3)
Analyze
the
combined
log.
Compare the
Combined log to
the Error Logs in
the Table in this
document.
Are any of the entries
contained in the
analyzed log?
No
Yes
Is "ACComm: Entering Alt-O2 Gas
Mode" in the analyzed log around
the Date and Time of the issue?
Follow the appropriate
No
Yes
Flowchart for the error listed.
Verify
Did Customer
with Customer
depress the
Button?
No
Yes
Check the Alt O2 Button and
wiring harness, ABS Filter
Normal Operation.
Connections, and ACB
No Service Issue.
Connections.
11
Flowchart
7-34
04/08 M1046983