Left-Side Cable Route; Front Side Of Display Connector Board And Lower Module Rack - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys

6.12.3 Left-side cable route

To access the left-side cable route, remove the following:
• the breathing system.
• the upper left-side cosmetic panel.
• the lower left-side cosmetic panel.
• the AGSS reservoir.
Route the cables through the cable clamps shown (
The upper left-side is a transition route for cabling:
• to the display arm (
).
G
• to the M-Gas chassis (
H
The extrusion includes an opening that accommodates the
ferrite bead (
) on power supply cable for the 15-inch
I
Anesthesia Monitor display.
6.12.4 Front side of Display Connector Board
and lower module rack.
Cables from the front side of the Display Connector Board and
the signal cable from the module rack route through an
opening in the left-side frame extrusion (
• System Power Interface to Display Unit (
• System Signal Interface to Display Unit (
• Airway Module (M-Gas) Power Supply Board (
• Signal cables to module rack (
To access the cable connectors, move the lower rack out of the
bay onto a suitable stand.
The power cable (
) to the module rack is routed from the right
O
side of the machine (TRM - Section 6.12.6).
Note: When replacing the module rack, ensure that the excess
cable length fall behind the rear drawer brace — away from the drawers.