GE Datex-Ohmeda Aisys Carestation Reference Manual page 204

Anesthesia machine
Table of Contents

Advertisement

Aisys
7-22
Test 8
Testing the bellows assembly
Note
If required, set up the Machine Test Tool and breathing system as shown in
Test 7.
1. Remove the bellows base manifold from the Bellows Module.
2. Insert appropriate test plugs into the bellows base manifold as shown to
the left.
Note: Position the bellows assembly so that the bellows remain collapsed
as you plug the ports.
Plug
3. Connect the tapered plug of the Machine Test Tool to the bellows base inlet
Plug
as shown to the left.
4. Position the bellows upright with the bellows collapsed.
Fresh Gas
5. Access the Gas Delivery Schematic (Section 12.3.2) of the Service
Application.
6. Set O
Flow to 0.2 l/min.
2
• Ensure that the Airway Pressure rises to ≥ 30 cm H
7. Set O
Flow to 0.00 l/min.
2
O.
2
04/08 M1046983