Factory defaults — Alarm limits
Alarm Limit
Value
Pmax High
40 cmH2O
(40 hPa, 4 kPa, 40 mbar,
30 mmHg)
MV High
10 l/min
MV Low
2 l/min
TV High
1000 ml
TV Low
Off
RR High
Off
RR Low
Off
Et CO2 High
8.0%
(60 mmHg or 8.0 kPa)
Et CO2 Low
3.0%
(23 mmHg or 2.0 kPa)
Fi CO2 High
Off
Fi O2High
Off
Fi O2 Low
21%
EtO2 High
Off
EtO2 Low
Off
Fi Iso High
5%
Fi Iso Low
Off
Et Iso High
Off
Et Iso Low
Off
Fi Sev High
8%
Fi Sev Low
Off
Et Sev High
Off
Et Sev Low
Off
Fi Des High
15%
Fi Des Low
Off
Et Des High
Off
Et Des Low
Off
Fi Enf High
5%
Fi Enf Low
Off
Et Enf High
Off
Et Enf Low
Off
Fi Hal High
5%
FI Hal Low
Off
Et Hal High
Off
Et Hal Low
Off