7.10.7 eVap Leak and Cassette troubleshooting

Cassette Related Error Log Entry
CASSETTE OVERFILL DETECTED
INVALID CASSETTE ID
VAP ENHANCED CASSETTE TEMPERATURE FAILURE
Is the Error entry
"Filling" related, "ID"
related, or "Temp"
related?
Filling
Filling Related
CASSETTE OVERFILL DETECTED
Isolate the cassette in use
during the event. Replace
the cassette if the issue
persists.
22
Flowchart
M1046983 04/08
Is the Error entry
Cassette
"Cassette" related or
"Leak" related?
Identification
Identification Related
INVALID CASSETTE ID
Normal for the Test Cassette
to cause this error entry.
If the issue occurs with
Therapy Cassettes, isolate the
issue to the cassette or the
ADB Board. Replace as
necessary.
Start
Leak Related Error Log Entry
INFLOW OUTFLOW CROSSCHECK FAILURE
INFLOW OUTFLOW CROSSCHECK FAILURE RECOVERABLE
Isolate the Aladin Cassette in use when the issue occurred.
Within the PC Service Application,
perform the Cassette Leak Test.
(Section 7.11)
Isolate the issue to a cassette or the eVap by
testing the therapy cassettes and the Test Cassette.
If a single cassette:
• Replace the cassette that leaks.
If the eVap:
• Replace the Valve Connector Assembly or O-ring
interface between the cassette and eVap.
Temperature
Temperature Related
ENHANCED CASSETTE TEMPERATURE FAILURE
VAP_ENHANCED_CASSETTE_TEMPERATURE_FAILURE
7 Troubleshooting
Leak
Isolate the cassette in use
during the event. Replace
the cassette if the issue
persists.
7-87