9.13.1 To remove
the Vent Engine
Access
Regulator Adjustment
A
M1046983 04/08
1. Disconnect pipeline supplies; close cylinder valves; bleed off pressure.
2. Remove the ABS breathing system.
3. Remove the Exhalation valve.
4. Remove the scavenging downtube.
5. Loosen the five captive screws (
Raise the cover to access the Vent Engine.
6. Disconnect the Vent Engine harness (
7. Disconnect the white tube-coupler (
transducer on the Ventilator Interface Board.
8. If present, disconnect the black tube-coupler (
indicator.
9. Disconnect the drive gas hose (
10. Loosen the three captive screws (
11. Lift the Vent Engine out of the housing.
12. To replace the Vent Engine, reassemble in reverse order.
) that hold the Vent Engine cover to the housing.
A
F
F
B
).
B
) — inline with tube to manifold pressure
C
), inline with tube to AGSS flow
D
).
E
) that hold the engine manifold to the housing.
F
9 Repair Procedures
C
F
E
D
9-37