Performance Test; Power Up Test; Key Test - GE Dash Responder Servicing Instructions

Table of Contents

Advertisement

GE Medical Systems
Information Technologies
Service Tool (2006861-001)
Oscilloscope
Defi paddles (217 333 01)
Internal Defi electrodes (217 308 01)
Electrode Cable for soft paddles (223 383 01)

Performance test

Power up test

Insert a charged battery.
Switch on the device by pressing the O
All LEDs should shortly flash once. (B
The indicator of the On-Off key is illuminated when the device is turned on.
Observe the display whether any error messages are displayed. The messages that are displayed
during a normal boot sequence are: S
C
GEMS IT year.
OPYRIGHT
After the boot sequence has been successfully performed the device starts up with a selected
energy according to the settings for the autosequence mode.

Key Test

Press both E
S
NERGY
K
T
.
EY
EST
Now press the following keys and verify the display entry:
Key
C
HARGE
S
YNC
left E
S
NERGY
ELECT
right E
S
NERGY
ELECT
left S
HOCK
S
RIGHT
HOCK
additional for Pacer Function
R
(
LEFT
ATE
PPM
R
(
RIGHT
ATE
PPM
O
(
LEFT
UTPUT
M
O
(
RIGHT
UTPUT
P
O
/O
ACER
N
FF
P
M
ACER
ODE
P
P
ACER
AUSE
N
: For the test of the pacer keys press the keys at least 2 seconds.
OTE
Switch the Dash Responder off and on.
Page 56 of 72
-O
N
FF
ATTERY
ELFTEST
keys simultaneously and then select S
ELECT
Display
Charge
Sync
Energy Select ∨
Energy Select ∧
Shock (left)
Shock (right)
Rate ∨
)
Rate ∧
)
Output ∨
A)
Output ∧
A)
M
Pacer On/Off
Pacer Mode
Pacer Pause
Servicing Manual Dash Responder V1.0
key.
C
LED may or may not flash)
HARGE
...; D
R
(software version);
ASH
ESPONDER
ERVICE
M
, C
T
and
ODE
OMP
EST
Revision F
2004768-001