M1046983 04/08
8. When replacing the sight glass (
down. Ensure that the two o-rings (
tightening the mounting screws.
9. When replacing the filler cap, thread the cord through the hole in the mask and tie a
knot so that
at least 95 mm
10.When replacing the label (
edge of the recess in the mask. Be careful to apply the correct label with regards to
whether the cassette has an enhanced temperature sensor or not.
11.When reattaching the mask, be sure that the contact retainer (
slot in the mask.
), orient it so that the rounded edge is facing
Item 10
) are seated in the sight glass before
Item 11
of cord remains external.
), be sure to align the top of the label with the top
Item 12
12
13
9 Repair Procedures
11
10
11
) engages the
Item 13
9-19