7.10.5 eVap Flow and Zero troubleshooting

Flow Related Error Log Entry
INFLOW CHECK VALVE FAILURE
OUTPUT FLOW LIMIT REACHED
CLOSED LOOP CONTROL FAILURE
Perform the following tests:
(Section 7.12)
• Backpressure Test
(Section 7.13)
• Inflow Check Valve Test
Replace any components that fail the tests.
20
Flowchart
M1046983 04/08
Is the Error entry
Flow
"Flow" related or
"Zero" related?
Start
Zero Related Error Log Entries
INFLOW ZERO POINT ERROR
OUTFLOW ZERO POINT ERROR
7 Troubleshooting
Zero
Within the PC Service Application,
verify the "zero" status
of the following fields:
• Inflow Flow
(Section 12.10.1)
• Outflow Flow
Replace the Flowmeter Subassembly.
7-85