GE Datex-Ohmeda Aisys Carestation Reference Manual page 205

Anesthesia machine
Table of Contents

Advertisement

Test 9
M1046983 04/08
Testing the flow sensor module, the circuit module, and the soda lime
canister
Plug
1. Separate the Bellows Module from the Circuit Module and re-install the Circuit/Flow
Sensor Module.
2. Connect short tubing between the inhalation and exhalation ports of the breathing
system.
3. Insert an appropriate test plug in the outlet port of the Circuit Module.
4. Access the Gas Delivery Schematic (Section 12.3.2) of the Service Application.
5. Set O
Flow to 0.2 l/min.
2
• Ensure that the Airway Pressure rises to ≥ 30 cm H
6. Set O
Flow to 0.00 l/min.
2
7. Remove the plug to release pressure.
7 Troubleshooting
O.
2
7-23