9.4 How to remove the tabletop
M1046983 04/08
The tabletop is held in place with four captive screws along the periphery of the pan
assembly (accessed from below the rim of the tabletop).
• Two screws (A) are at the front of the tabletop: one screw is at the right corner of the
tabletop, one is near the O
• To access the remaining two screws (B), you must remove the ABS: one screw is at the
left corner of the tabletop, one is near the APL Valve.
B
B
Flush button.
2
A
9 Repair Procedures
A
9-7
Need help?
Do you have a question about the Datex-Ohmeda Aisys Carestation and is the answer not in the manual?