GE Datex-Ohmeda Aisys Carestation Reference Manual page 210

Anesthesia machine
Table of Contents

Advertisement

Aisys
Test 15
Note: To ensure a air-tight
seal, use the corresponding
plug as illustrated for the
original flow sensor ( A )
or the new,
moisture resistant
(offset) flow sensor ( B ).
7-28
Testing a flow sensor including the Ventilator Monitoring Assembly and
interfacing components
1. Remove Flow Sensors from the Flow Sensor Module.
2. Attach the Flow Sensor to the bulkhead connector.
3. Plug each Flow Sensor as shown.
4. Connect the low-pressure leak test device to the open end of the Flow Sensor.
5. Compress and release the bulb until it is empty.
6. If the bulb inflates in 30 seconds or less, there is a leak. The leak may be through the
connector o-rings, in the internal tubing, or in the Transducer on the VIB.
A
B
04/08 M1046983