GE Datex-Ohmeda Aisys Carestation Reference Manual page 276

Anesthesia machine
Table of Contents

Advertisement

Aisys
Error Log Entry
VAPCHK INCHECK LEAK
VAPCHK INFLOW LEAK
VAPCHK INFLOW
OUTFLOW
VAPCHK INFLOW ZERO
VAPCHK MAX FLOW
7-94
Condition / Test Configuration
Inflow Check Valve leak exceeds limit.
Test:
Mixer Flow: 8 l/min
Inflow Valve: On
Scav Valve: Off
Outflow Valve: Off
Mixer Flow reduced to 2 l/min.
Monitor eVap Inflow Flow for negative flow.
Inflow Valve leak exceeds limit.
Test:
Mixer Flow:2 l/min
Inflow Valve: Off
Scav Valve: On
Outflow Valve: On
Prop Valve: 65534
Inflow flow values indicate flow equal to Inflow
Valve leak rate.
Input and Output Flow Meters do not match to
within limit
Test:
Mixer Flow: 0.5 l/min
Inflow Valve: On
Scav Valve: Off
Outflow Valve: On
Prop Valve: 65534
Inflow Meter Zero Valve fails to operate
Test:
Mixer Flow: 0.5 l/min
Inflow Valve: On
Scav Valve: Off
Outflow Valve: On
Prop Valve: 65534
Verify Inflow reads at or near 0.5 l/min
Uncheck Zero Valves Auto and select Inflow
Zero – On. Verify LED comes on, valve clicks,
and Inflow reads at or near 0 ml.
Unable to achieve required maximum eVap
output flow
Action/Troubleshooting
Replace:
• Inflow Check Valve
Replace:
• Inflow Valve (right side valve block assembly
as viewed from front).
Replace:
• Flow Meter Block Assembly
• Agent Delivery Board
Replace:
If the test fails:
• Input Flow Meter Zero Valve (right side of Flow
Meter Block Assembly as viewed from the
front)
• Agent Delivery Board
If the test passes:
• Flow Meter Block Assembly
Replace:
• Backpressure Valve
• Proportional Valve
• Flow Meter Block Assembly
• Agent Delivery Board
04/08 M1046983