10.18.2 Suction Control Module

8
7**
only with Venturi
7c
7b
7a
Item Description
1
Gauge, 760 mmHg
1a
O-ring, Gauge (included with gauge assy, 2ea. required)
2
Control panel assembly, with suction regulator knob and mode control knob
3
Regulator Module (plugs into manifold assembly)
3a
O-ring, Regulator Module, Large (included with regulator module)
3b
O-ring, Regulator Module, Stem (included with regulator module)
4
Manifold Assembly, without Gauge and Regulator Module
5
Screw, #6 - 2 inch
6
Filter
7
Pilot valve adapter assembly (includes plunger, jam nut, and valve assembly)
8
Cap, white
* Lubricate the regulator module o-rings and the mating bore of the manifold sparingly with Dow 111 lubricant.
** Drop the plunger (7a), round end first, into the manifold. Thread the pilot valve into the manifold body. Set the
mode switch to raise the plunger. Adjust the pilot valve (7b) so that the plunger actuates the pilot valve
approximately half of its travel. Tighten the jam nut (7c).
M1046983 04/08
6
*
5
4
3b*
3a*
3
2
10 Illustrated Parts
1a
1
Stock Number
1009-3227-000
6700-0133-500
1009-3274-000
6700-1225-800
6700-0136-500
0210-0527-300
1009-3277-000
1009-3340-000
0206-5159-300
1009-3278-000
1009-3385-000
10-27