12.8.5 Gas Delivery Subsystem
M1046983 04/08
Mixer Post/Checkout Test Results
Mixer Post/Checkout Test Results
Label
Alt O2 Valve Leak
O2 Valve Leak
Balance Gas Valve Leak (Mixer)
O2 Flow Test
Balance Flow Test
12 Service Application
Test Results
Pass
not performed (if O2 Valve Leak fails)
Not done. No supply pressure
Not done. Selector valve incorrect state
Fail. Selector valve leaks
Fail. Proportional valve leaks
Pass
Fail. O2 Bypass valve leaks
Pass
Not done. No supply pressure
Not done. Selector valve incorrect state
Fail. Selector valve leaks
Fail. Proportional valve leaks
Pass
Not done. No supply pressure
Not done. Selector valve incorrect state
Fail, 3 LPM test failed
Fail, 10 LPM test failed
Pass
Not done. No supply pressure
Not done. Selector valve incorrect state
Fail, 3 LPM test failed
Fail, 10 LPM test failed
12-23