9.25 Display arm adjustments

9.25.1 Display arm
counterbalance
adjustment
9-58
Adjustments can be made to the following components to maintain proper positioning
of the displays:
• Counterbalance of the display arm (up and down movement)
• Wrist casting (side to side movement of displays)
• Clutch bearings (tilt position of the displays)
1. Remove the display arm cover by loosening the five captive screws.
2. Ensure that the displays are not positioned over the top shelf.
3. Access the counterbalance adjustment nut at the back of the display arm; it is
approximately 6.5 cm inside the metal frame of the arm.
4. Using a 1/2-inch socket and at least a 8-cm extension, turn the adjustment nut
clockwise to add more counterbalance or counterclockwise to remove
counterbalance.
5. Adjust the counterbalance to the desired setting.
• The display arm should move up or down freely when assisted but remain in
position after moving it to the desired vertical position.
• The arm should not fall or rise independently when the counterbalance is correctly
adjusted.
6. Reinstall the display arm cover.
04/08 M1046983