GE Datex-Ohmeda Aisys Carestation Reference Manual page 314

Anesthesia machine
Table of Contents

Advertisement

Aisys
non-DES cassette
non-DES cassette
ZIF
9-18
6. To replace the Agent Cassette board (
• For a non-DES cassette,
use a2.5-mm hex wrench to remove the mounting screws (
• For a DES cassette,
use a 5.5-mm socket to remove the fully exposed mounting screw (
To remove the screw under the filler assembly (
wrench or a small size adjustable open end wrench to loosen the screw to the
point where you can spin it out with your fingers. When mounting the board,
ensure that you include the insulating washer under the screw head to prevent
board damage when tightening this fastener. Place the hardware into the board
mounting hole prior to positioning the board on the DES cassette.
6
7
7. Note that the ribbon cable attached to the Agent Cassette board uses a
insertion force) connector; pull tabs toward ribbon cable to release (push down after
inserting cable). The conductors on the ribbon cable face the top of cassette.
• Cassettes without enhanced temperature sensing do not include this ribbon
cable.
) remove the two mounting screws.
Item 6
Item 9
DES cassette
8
DES cassette
).
Item 7
).
Item 8
), use a 5.5-mm open end
6
9
(zero
ZIF
ZIF
04/08 M1046983