Error Log Entry
Action/Troubleshooting
VENT +12.5V FAIL
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and +12.5 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT +6V FAIL
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and +6.0Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT 1.22V FAIL
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and 1.22 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT -6V FAIL
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT ADC VREF FAIL
In the Service Software / Vent Interface Bd Power Diagnosis, view the "Vent Int Bd 10VA Voltage" from
Board Supplies:
If "Vent Int Bd 10VA Voltage" reads "OK, and 3.2 Vdc reads "Fail", replace the VIB.
If "Vent Int Bd 10VA Voltage" reads "Fail" and the "Vent Int Bd 10VA Voltage" from the Anes Cntrl Bd
reads "OK", Check cabling between ACB and VIB.
VENT AIRWAY
OVERPRESS SIGNAL
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
VENT AIRWAY
OVERPRESS SIGNAL
FAIL
No Service Action.
Reboot system. If problem continues, check Airway Pressure signal in Service Mode.
M1046983 04/08
Alarm Text
Condition (Basic info)
Vent Fail. Monitoring
Nominal 12.5V <11.3 Vdc or
Only
>13.13 Vdc
Vent Fail. Monitoring
VSIB +6V out of range<5.51
Only
Vdc or > 6.5 Vdc
Vent Fail. Monitoring
Voltage < 1.074Vdc or Voltage
Only
> 1.367 Vdc
Vent Fail. Monitoring
VSIB -6V out of range<-6.72
Only
Vdc or > -5.28 Vdc
Vent Fail. Monitoring
VSIB ADC3.200V ref voltage
Only
out of range <3.179 or
>3.221 Vdc
Inspiration stopped
High airway overpressure
signal set.
Vent Fail. Monitoring
Ventilator SIB indicates the
Only
High Airway overpressure
signal was set and Paw < 90
cmH
cm H
Priority Source Enabling Criteria
Medium
Medium
-6.0Vdc reads "Fail", replace the VIB.
Medium
Medium
O and Pmanifold <80
2
O.
2
7 Troubleshooting
AC,
Vent
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
AC,
Vent +12.5 V (10 VA) is
Vent
OK
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
High
AC,
Vent +12.5 V (10 VA) is
Vent
OK
AC,
Mechanical Ventilation
Vent
On
AC,
Mechanical Ventilation
Vent
On