9.6 How to access dashboard components

w
CAUTION
9-10
1. Bleed all gas pressure from the machine (Section 9.2).
2. Ensure that all cylinder and pipeline pressures read zero before proceeding.
3. Remove the tabletop (Section 9.4).
4. Remove the upper bezel located above the dashboard.
5. Loosen 10 captive dashboard mounting screws and move the dashboard forward to
the stop position.
When replacing the dashboard back into the machine, take care not to trap, kink,
or snag any tubing or wiring harnesses.