9.7 Replace electronic vaporizer and components

M1046983 04/08
1. Move the dashboard forward to the service position (Section 9.6).
2. Lift the assembly slightly to release the hanging pin from the rail.
(A)
3. Position the dashboard so that the eVap support bracket (
chassis and the bottom edge of the dashboard (
enclosure.
4. In the forward position, the following eVap components can be replaced without
removing the eVap from the machine. Refer to the following sections for details.
• Flowmeter subassembly
• Flowmeter inflow and outflow zero valves
• Manifold temperature sensor board
• Agent delivery board
• Cassette interface board
5. To replace the eVap, or components not mentioned above, remove the eVap from
the machine as detailed in the following section.
9 Repair Procedures
(B)
) rests on the edge of the
A
) straddles the pan electronic
B
9-11