9.21 Replace ABS breathing system components

9.21.1 Replace
Bag/Vent switch
assembly
9-52
1. Remove the ABS breathing system.
2. From the underside, remove the bellows base manifold (A) and fully loosen the two
captive screws (B) at the bag port side of the APL/BTV manifold.
3. From the topside, rotate the Bag/Vent switch cartridge counterclockwise until the
Bag/Vent switch outlet port (C) clears the bellows housing.
4. Lift out the Bag/Vent switch cartridge from the housing.
5. Replace the Bag/Vent switch cartridge in reverse order.
6. Reinstall the ABS breathing system.
7. Perform the checkout procedure (Section 3).