GE Datex-Ohmeda Aisys Carestation Reference Manual page 199

Anesthesia machine
Table of Contents

Advertisement

Test 3
M1046983 04/08
Testing the airway pressure transducer, and Port 1 and Port 3 u-cup seals
1. Access the Gas Delivery Schematic (Section 12.3.2) of the Service Application.
2. Set O
Flow to 0.1 l/min.
2
3. Attach the Machine Test Tool to the breathing system interface ports (using the
alignment post) as shown above.
4. Occlude the tapered plug.
• the Airway Pressure reading should increase.
• If not, there is a leak in the tested circuit.
5. Set O
Flow to 0.00 l/min.
2
7 Troubleshooting
Alignment
Post
7-17