Error Log Entry
Action/Troubleshooting
ADB 10VA POWER
ERROR
Disconnect ADB power harness and restart the system.
If the error does not reappear in the log,
• replace the ADB and retest.
If the error persists,
• proceed to the interconnect fault isolation procedure (Section 7.10.12).
(Also see Section 7.10.)
ADB VOLTAGE ERROR
Replace ADB and retest.
(Also see Section 7.10.)
AGENT LEVEL LOST
Check, replace, and retest as needed in the following order:·
• All four magnets on Cassette Interface Board fingers draw down and make contact with Test Cassette
• Troubleshoot/Test cassette(s)
• Cassette Interface Board to Agent Delivery Board cable connections are secure
• Cassette Interface Board
• Agent Delivery Board
AGENT LEVEL OVER
RANGE
Check, replace, and retest as needed in the following order:·
• All four magnets on Cassette Interface Board fingers draw down and make contact with Test Cassette
• Troubleshoot/Test cassette(s)
• Cassette Interface Board to Agent Delivery Board cable connections are secure
• Cassette Interface Board
• Agent Delivery Board
M1046983 04/08
Alarm Text
Condition (Basic info)
Vaporizer Failure
Overcurrent condition
detected by the AC. Circuit
disabled.
Vaporizer Failure
One or more of the measured
ADB voltages have failed.
These include the ADC
reference voltage, 12P1
power supply, and five volt
supply.
Check agent level
Agent level reporting changes
from being reported to agent
level no longer available.
contacts
Analog agent level (Aladin1
Desflurane) reading was valid
but then went out of range
(high).
contacts
7 Troubleshooting
Priority Source Enabling Criteria
AC
AC
Low
AC
Cassette inserted And
Agent level is reported
And Agent ID is known
(agent ID is not none).
7-47