Display Arm - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys

10.42 Display arm

1
11
9 (10)
8
**
5
6
12
13 (14)
10-80
2
3*
4
Item
Description
1
Display arm assembly
2
Label, arm cable routing
***
3*
Washer, friction shoulder
4
Bearing, Nyliner with key
5**
Spacer, split plastic
6
Support, shaft bottom
7
Screw, M6x16 Sems
7
8
Support, shaft top
9
Screw, M6x16 SKT HD CAP
10
Lockwasher, M6 internal
**
11
Roll pin, 0.25 OD
12
Washer, shoulder
13
Screw, M4x12 SKT HD CAP
14
Lockwasher, M4 internal
* Lubricate both sides of friction washer (
the first 3 cm of shaft next to washer.
** Ensure end-gap of spacer is opposite slot in lower support.
*** Push spacer (
Item 5
Note:
When replacing the display arm, loosen the mounting screws (
the bottom shaft support (
before replacing the shoulder washer (
Stock Number
1011-3290-000
1011-3568-000
1011-3390-000
1006-3228-000
1011-3393-000
1011-3388-000
0144-2436-109
1011-3389-000
1011-3894-000
0144-1118-130
0201-0757-300
1407-3814-000
1102-3006-000
0144-1118-128
) and completely around
Item 3
) into lower support using shoulder on shaft.
) to ease alignment. Retighten the screws
Item 6
).
Item 12
Qty
2
4
4
) for
Item 7
04/08 M1046983