GE Datex-Ohmeda Aisys Carestation Reference Manual page 84

Anesthesia machine
Table of Contents

Advertisement

Aisys
Mechanical expiration
(EZchange and
condenser
ON)
D
6
6
AP Airway Pressure
D
Drive gas
FG Fresh Gas
SGR Sample Gas Return
4
Expiratory flow
5
Flow to bellows
6
To scavenging
Figure 2-39 • Mechanical expiration through Condenser with EZchange Canister and Condenser ON
2-52
Drive-gas flow stops and the exhalation valve opens. Exhaled gas flows from the
patient (4), through a unidirectional valve (expiratory check valve) and into the bellows
(5). Residual drive gas (D) flows out of the bellows to the scavenging system (6).
If PEEP is selected, static pressure on the pilot port of the exhalation valve sets the
PEEP level.
During exhalation, fresh gas flows backwards through the Condenser and absorber
(FG) into the expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system, the
returned gas (SGR) enters the breathing system after the expiratory check valve.
5
5
AP
FG
SGR
FG
FG
4
04/08 M1046983

Advertisement

Table of Contents
loading

Table of Contents