Anesthesia Control Board - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys
9.12.3 Anesthesia
Control board
9-34
To replace the Anesthesia Control board, first remove the Power Supply/Power
Controller board assembly (Section 9.12.2). Then, follow the procedure below:
1. Disconnect the cables coming from the following components:
• the large ribbon cable from the Pan Connector board (A),
• the harness from the Display Connector board (B),
• the small ribbon cable from the Display Connector board (C),
• the harness from the Power Controller board (D),
• the harness from the Pan Connector board (E).
A
E
D
2. Loosen the six screws (circled) that hold the Anesthesia Control board to the
enclosure.
3. Lift the Anesthesia Control board slightly to release it from the keyhole slots.
4. To replace the Anesthesia Control board, reassemble in reverse order.
5. Load Software.
6. Check/Reconfigure the Machine Configurations (Install/Setups).
7. Perform User Calibrations (Section 4.5.2).
8. Zero Gas Transducers (Section 4.5.8).
9. Perform all Ventilator Calibrations (Section 5.4).
B
C
04/08 M1046983