4.5.2 User Calibration
4-20
menu
Menu Item
Flow and Pressure
Circuit O2 Cell
Airway Gas
Backlight Test
Previous Menu
Message text
Calibrate the flow and pressure sensors.
Remove flow sensor module to start.
Replace when Pass or Fail message appears.
Calibrate Circuit O2 Cell.
21% O2
Remove Flow Sensor Module.
Expose O2 Cell to Room Air.
Start 21% calibration.
May take 3 min.
100% O2
1. Reconnect flow sensor module.
2. Set Bag/Vent switch to Vent.
3. Select 100% O2.
Start Gas Calibration. Calibrate CO2, O2, N2O, and
agent measurements. (MGAS module must be installed).
Gas calibration is not available during gas sampling
warm-up and certain alarms.
Push ComWheel to test back lights. Test every month.
This test turns off one backlight to test the other light.
Screen brightness may change during test.
Return to the previous menu.