GE Datex-Ohmeda Aisys Carestation Reference Manual page 217

Anesthesia machine
Table of Contents

Advertisement

Anesthesia Control Board Troubleshooting
Upgrade to current
Software Version.
(Section 8.2)
No
12
Flowchart
M1046983 04/08
Start
Yes
Is Software Version current?
No
No
Does ACB communicate in
Service Application?
Yes
Can the error be duplicated?
Yes
Replace ACB
7 Troubleshooting
Error Entry
ACB CLOCK SPEED
ACB CPU TEST FAIL
ACB EEPROM FAIL
ACB FLASH FAIL
ACB HW WATCHDOG
ACB MICROPROC ERROR
ACB RAM ERROR
ACB REDUNDANT MEMORY FAIL
ACB SW ERROR
ACB SW WATCHDOG
ACB UNEXPECTED RESET
ACB RAM MEMORY TEST FAILURE
LOSS OF GAS DELIVERY USER SETTINGS
LOSS OF VENT DELIVERY USER SETTINGS
No
Contact
Technical Support
7-35