GE Datex-Ohmeda Aisys Carestation Reference Manual page 227

Anesthesia machine
Table of Contents

Advertisement

Error Log Entry
Action/Troubleshooting
+5V H AMP GAS
SUPPLY XDUCERS
Reboot system. If problem continues:
1. With system in Standby, disconnect all gas supply pressure transducers from ABS Filter board.
2. Reboot system.
• If problem continues, replace the ACB.
• If error is no longer present, set system to Standby and reconnect one pressure transducer at a time.
3. Reboot system and check for error with each transducer connected. Replace the transducer that
causes the error to appear.
12 HR TEST
At next available time, move the system switch from the On position to the Off position, then back to the
On position.
ACB 4.096V ADC REF
Reboot System. If problem continues, replace the ACB.
ACB ADC FAIL
Reboot System. If problem continues, replace the ACB.
ACB CLOCK SPEED
Reboot System. If problem continues, replace the ACB.
ACB COM FAIL
Reboot System. If problem continues, replace the ACB.
ACB CPU TEST FAIL
Reboot System. If problem continues, replace the ACB.
ACB DCB COM FAIL
Reboot System. If problem continues, check the ACB to Mixer communication LED's (or VIB
communication LED's). If the RCV and XMT (or TXD and RXD) LED's indicate activity, check DU cable
connections, replace Display Controller PCB if problem continues. If the RCV and XMT (or TXD and RXD)
LED's indicate no activity, check the Anesthesia Control board connection, replace the Anesthesia
Control board if problem continues.
M1046983 04/08
Alarm Text
Condition (Basic info)
Cannot read gas
Status bit shows current high. Medium
supply pressures
Turn power Off and On
System has been operating
for self tests
for longer than 12 hours
without a power up self test.
Cannot monitor gas
<4.018 or > 4.176 Vdc
supplies
Cannot monitor gas
ADC timeout on any MUX
supplies
channel.
System Malfunction
AC Clock frequency
> 1.1* (expected value) or
< 0.9* (expected value).
AC Clock frequency incorrect.
System Malfunction
After establishing initial
communication, the DC does
not receive any messages
from AC in 10 sec.
System Malfunction
CPU instruction Test Failure
System Malfunction
The Anesthesia Computer
receives no system state
messages from the Display
Computer for 10 seconds.
7 Troubleshooting
Priority Source Enabling Criteria
AC
Pressure transducer
+10VA turned On.
Low
AC -
System state is in
Vent
Checkout.
DC
checks
enable
criteria
Low
AC
Low
AC
High
AC
High
DC
High
AC
High
AC
Initial communications
established.
7-45