GE Datex-Ohmeda Aisys Carestation Reference Manual page 242

Anesthesia machine
Table of Contents

Advertisement

Aisys
Error Log Entry
Action/Troubleshooting
MAN CASS OVER
UNDER TEMP
Operating temperature as measured by one of the Electronic Vaporizer temperature sensors was out of
allowed operating range.
MANIFOLD PAW
SENSOR FAIL
In the Service Software / Vent Flow and Pressure Diagnostics, verify the Manifold Flow counts is
800 ± 250.
Disconnect the White in-line connector in the Manifold Pressure. If the counts return within specified
range, check for occlusions in the Bulkhead harness.
If the counts do not return within the specified range, replace the VIB.
MANIFOLD
PRESSURE SENSR
FAILURE
In the Service Software / Vent Flow & Pressure Diagnostics, verify the Manifold Flow counts is
800 ± 250.
Disconnect the White in-line connector in the Manifold Pressure. If the counts return within specified
range, check for occlusions in the Bulkhead harness.
If the counts do not return within the specified range, replace the VIB.
MANIFOLD SENSOR
CAL ERROR
Could be caused by bad span calibration or leaky vent Insp Flow Valve.
MGAS CHECK SAMPLE
GAS OUT >20 SEC
Replace sample line. See AM TRM for further Troubleshooting.
MGAS INLET FILTER
RESIDUE >40 SEC
Replace D-Fend. See AM TRM for further Troubleshooting.
MGAS LINE BLOCKED
>20 SEC
Replace sample line. See AM TRM for further Troubleshooting.
7-60
Alarm Text
Condition (Basic info)
Check cassette. Set
Either manifold temperature
agent.
reading or cassette
temperature reading outside
of limit.
Vent Fail. Monitoring
Calibration failure at bootup.
Only
Indicates a calibration
failure at bootup.
Indicates the Manifold
Pressure zero failed.
Check sample gas out
MGAS SPEC. Continuous
Occlusion Bit set.
Replace D-Fend
MGAS SPEC (Residue build-
up on the water trap
membrane. This decreases air
flow).
Sample line blocked
MGAS SPEC states The
sample tubing inside or
outside the monitor blocked,
or the water trap is occluded.
Priority Source Enabling Criteria
Medium
AC
De-
escalating
Medium
AC,
Vent
ACB
ACB
Medium MGAS MGAS present and
MGAS communicates
continuous occlusion
for 20 seconds
Medium MGAS MGAS present and
MGAS communicates
this the Replace Trap
alarm bit for 40
seconds
Medium MGAS MGAS present and
MGAS communicates
this the continuous
occlusion alarm for 20
seconds
04/08 M1046983