4.2.4 Factory Defaults

Parameter
Value
Vent Mode
VCV
TV (tidal volume)
500 ml
Pinsp
5 cmH2O
(5 hPa, 0.5 kPa, 5 mbar, 4.0 mmHg)
RR
12 /min
Mech RR
12 /min
Tinsp
1.70
I:E
1:2.0
Trig. Window
25%
Flow Trig.
2 l/min
End Breath
30%
Psupport
2 cmH2O
Tpause
Off
Backup Time
30 s
Total Gas Flow
6 l/min with Circle and
10 l/min with Non-Circle
O2 Flow l/min
6 l/min with Circle and
10 l/min with Non-Circle
Control Style used for
O2%
default setting
Flow
Balance Gas
Air
Circuit
Circle
Gas Outlet installed
SCGO
Paw Color
Yellow
Flow Color
Green
Resp Color
White
CO2 Color
White
Paw Units
cmH2O
Weight Units
kg
Temperature Units
C
4-8
The following table lists the factory defaults for parameters. The table on the
next page lists the factory defaults for alarm limits.
Factory defaults — Parameters
when Install/Service Fresh Gas
Controls menu item is set to
"O2% or "User."
when Install/Service Fresh Gas
Controls menu item is set to
"Flow."
Parameter
Value
CO2 Units
%
Altitude
300 m
Gas Supply Pressure
kPa
Units
Decimal marker
. <dot>
Language
English
Gas Supply Colors
ISO
(O2 = white, N2O = blue,
Air = blk/wht)
O2 flow tube
Right side
Vent drive gas
O2
Rise Rate with Auto
Auto
(PCV, SIMV/PSV,
PSVPro, SIMV-PC
Modes)
Rise Rate without
5
Auto (PCV-VG Mode)
N2O enabled
No
Spirometry Source
Vent
Patient/Sensor Type
Adult
Auto MV Limit
Off
Alarm Volume
3
MV/TV Alarms
On
Volume Apnea
On
Sweep speed
Fast
Split screen
Gas
Wave field 1
Paw
Wave field 2
Flow
Wave field 3
CO2
04/08 M1046983