Fresh Gas And O 2 Flush Flow (With Scgo) - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys
2.12.4 Fresh gas and O
To ABS (Circle)
breathing system
ABS
Figure 2-43 • Fresh gas and O
2-56
flush flow (with SCGO)
2
Fresh gas (1) flows from the electronic vaporizer (EV) to the SCGO assembly.
With the Circle system selected, fresh gas flow is channeled to Port 3 of the breathing
system (before the inspiratory check valve).
The output of the O
Flush regulator (2) is channeled to the O
2
When activate, O
flush flow joins the fresh gas flow in the SCGO assembly.
2
5.4 psi
O2 Flush
2
Switch
Circle or CGO
Selector Switch
(SCGO)
1
SCGO
Variant
flush flow (to ABS)
2
To
Scavenging
EV
From
Mixer
1
SCGO
Assembly
Absorber
O2
Sensor
3
Insp Flow
2
1
Pilot
Paw Trans
150 cm H2O
FG Pressure
Limiting Valve
Flush valve.
2
From
O
Supply
2
To
Mixer
2
04/08 M1046983

Advertisement

Table of Contents
loading

Table of Contents