Error Log Entry
Action/Troubleshooting
INFLOW ZERO 10VA
POWER ERROR
If the P3 indicator never lights or lights only briefly,
• proceed to the interconnect fault isolation procedure (Section 7.10.12).
If the P3 indicator remains lit for more than 2 seconds,
• isolate the failed valve (Inflow, Inflow zero or Outflow zero) circuit with manual valve controls.
Replace and retest in the following order (Also see Section 7.10.):
• Valve
• ADB
INFLOW ZERO POINT
ERROR
Replace and retest in the following order (Also see Section 7.10.):
• Inflow Zero Valve
• Flowmeter Block
• ADB
INSP FLOW SENSOR
EEPROM FAILURE
Replace the Inspiratory Flow Sensor.
INSERT CASSETTE
Occurs whenever a cassette is not inserted.
INVALID CASSETTE ID
If failure occurs with multiple cassettes of the same type (agent),
• insert test cassette and verify test cassette is identified in the PC Service App.
Remove test cassette and verify PC Service App indicates 'NONE'.
If either PC Service App test fails,
• replace ADB.
(Also see Section 7.10.)
7-58
Alarm Text
Condition (Basic info)
Vaporizer Failure
Overcurrent condition
detected by the AC. Circuit
disabled.
Vaporizer Failure
Input flowmeter measured
value during zeroing is out of
range.
Replace insp flow
EEPROM cal data read failure
sensor
The system does not detect a
vaporizer cassette
Cannot identify
Invalid cassette ID code
cassette
Priority Source Enabling Criteria
AC
AC
Low
AC,
Vent
Low
AC
Agent delivery not on
(agent Off or state is
checkout)
Medium
AC
De-
escalating
04/08 M1046983