Aisys

7.10.10 eVap Temperature troubleshooting

Cass Temp
Cassette Temperature Related
CONDENSATION CONDITIONS EXIST
VAP_CONDENSATION_CONDITION_EXISTS
Indicates a warm
cassette was installed.
No service action required.
25
Flowchart
7-90
Start
Is the Error entry
"Cass Temp",
"Cass x-ducer",
or "Manifold"
related?
Cass x-ducer
Cassette x-ducer Related
CASSETTE TEMPERATURE FAILURE
CASSETTE TEMPERATURE EEPROM FAILURE
Replace
• Cassette Temperature Sensor.
• the ADB Board
Error Log Entry
CASSETTE TEMPERATURE EEPROM FAILURE
CASSETTE TEMPERATURE FAILURE
CONDENSATION CONDITIONS EXIST
MAN CASS OVER UNDER TEMP
MANIFOLD TEMPERATURE EEPROM FAILURE
MANIFOLD TEMPERATURE FAILURE
VAP CASS TEMP SENSOR COMPAT FAIL
VAP CONDENSATION CONDITION EXISTS
Manifold
Manifold Temperature Related
MANIFOLD TEMPERATURE EEPROM FAILURE
MANIFOLD TEMPERATURE FAILURE
MAN CASS OVER UNDER TEMP
Check connection of the temperature
sensors to the ADB.
If the problem persists,
• check the value of the temp sensors using
the PC Service App.
(Section 12.10.1)
If reading does not change when the temp
cable is disconnected,
• replace the ADB.
If the reading does change,
• replace the temp sensor out of range.