6.11.3 Left-side cable route

To access the left-side cable route, remove the
following:
• the breathing system.
• the upper left-side cosmetic panel.
• the lower left-side cosmetic panel.
• the AGSS reservoir.
Route the cables through the cable clamps shown
(
).
arrows

6.11.4 Upper left-side cable route

The upper left-side is a transition route for cabling.
• to the display arm (
).
G
• to the M-Gas chassis and the upper module rack (
The extrusion includes an opening that accommodates the ferrite
bead (
) on the power supply cable for the 15-inch Anesthesia
I
Monitor display.

6.11.5 Front side of Display Connector Board

Cables from the front side of the Display Connector Board route
through an opening in the left-side frame extrusion (
• System Power Interface to Display Unit (
• System Signal Interface to Display Unit (
• Airway Module (M-Gas) Power Supply Board (
To access the cable connectors, you need to remove one of the
following components depending on machine configuration:
• a drawer.
• the cassette storage bay assembly.
K
J
6-18
).