GE Datex-Ohmeda Aisys Carestation Reference Manual page 256

Anesthesia machine
Table of Contents

Advertisement

Aisys
Error Log Entry
Action/Troubleshooting
VAPORIZER
SUBSYSTEM COMM
FAILURE
Verify harnesses from ACB to PCB and ACB to ADB are connected.
Replace and retest, in the following order (Also see Section 7.10.):
• ADB
• ACB
• PCB/Harnesses
VAP SENSOR ERROR
Check all temp/pressure/flow sensors connections to the ADB.
Cycle power.
If problem persists,
• run PC Service App Vaporizer Test (Section 12.10.2).
(Also see Section 7.10.)
Vent check stage 1
failed.
Checkout failed.
Check for other errors in the error logs.
Vent check stage 1
fails.
Checkout failed.
Check for other errors in the error logs.
Vent check stage 2
failed.
Checkout failed.
Check for other errors in the error logs.
Vent check stage 2
fails.
Checkout failed.
Check for other errors in the error logs.
VENT FLOW VALVE FAIL
DAC
Reboot System. If problem continues, in the Service Software / Vent Flow and Pressure Diagnosis,
increase the Flow Valve counts and view the Flow Valve Feedback mV and Counts. Verify the settings
match.
7-74
Alarm Text
Condition (Basic info)
Vaporizer Failure
Anesthesia Computer and
Agent Delivery Subsystem
communication lost or error
for greater than 1 second.
Vaporizer Failure
One or more of the vaporizer
sensors is grossly out of range
(indicating electrical fault or
disconnect).
A message or failure
displayed during the system
checkout.
A message or failure
displayed during the system
checkout.
A message or failure
displayed during the system
checkout.
A message or failure
displayed during the system
checkout.
Vent Fail. Monitoring
Incorrect DAC feedback for 3
Only
consecutive readings
Priority Source Enabling Criteria
AC
AC
DU
DU
DU
DU
Medium
AC,
Vent
04/08 M1046983