2.8 Ventilator Interface board

Figure 2-17 • Ventilator Interface board
2-22
The Ventilator Interface board (A) provides the electrical and/or pneumatic interface to
the following:
• Inspiratory (B) and expiratory (C) flow sensors (transducers)
• Patient airway (D) and manifold (E) pressures (transducers)
• Oxygen sensor (in breathing system)
• Inspiratory Flow Valve
• Gas Inlet Valve
• SCGO solenoid, SCGO/CGO position switches (if SCGO installed)
• Canister release switch
• ABS On switch
• ACGO position switch (if ACGO installed)
• O
Flush switch
2
• Bag/Vent switch
• Accessory Power (for task lights)
The Ventilator Interface board functions are managed locally by a microcontroller. The
microcontroller communicates data values to the controlling CPU via an RS-422 serial
interface.
B
D
A
E
C
04/08 M1046983