Storz 11272 V Series Instruction Manual page 71

Fexible cmos video-cysto-urethroscope
Hide thumbs Also See for 11272 V Series:
Table of Contents

Advertisement

The flexible CMOS Video-Cysto-Urethoscope 11272 V/VU in conjunction with Monitor 8402 ZX is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum out-
put power of trans-
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
69
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the
KARL STORZ flexible CMOS Video-Cysto-Urethoscopes 11272 V/VU in conjunction with Monitor 8402 ZX
mitter
150 kHz to 80 MHz
[W]
d=[3,5/3]P
0.01
0.1
0,1
0.4
1
1.2
10
3.7
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 206
Separation distance d [m]
according to frequency of transmitter
80 MHz to 800 MHz
d=[3,5/3]ÖP
0.1
0.4
1.2
3.7
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
800 MHz to 2.5 GHz
d=[7/3]P
0.2
0.7
2.3
7.4

Advertisement

Table of Contents
loading

This manual is also suitable for:

11272 vu series11272 vk series11272 vuk series8402 zx