Preparing the Patient

Resting ECG Electrodes

Standard 12 Lead Placement
3-4
To acquire a standard 12 lead ECG, use the placement shown in the
following illustration.
AHA
IEC Label
Label
A
V1 red
C1 red
B
V2 yellow
C2 yellow
C
V3 green
C3 green
D
V4 blue
C4 brown
E
V5 orange
C5 black
F
V6 purple
C6 purple
G
LA black
L yellow
H
LL red
F green
MAC™ 1600
Electrode Placement
Fourth intercostal space at the right sternal border.
Fourth intercostal space at the left sternal border.
Midway between location B and D.
Mid-clavicular line in the fifth intercostal space.
Anterior axillary line on the same horizontal level as
D.
Mid-axillary line on the same horizontal level as D
and E.
Left deltoid.
Above left ankle. (Alternate placement, upper leg as
close to torso as possible.)
026A