Query/Retrieve Setup - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

System Setup

3-9-7 Query/Retrieve Setup

NOTE:
3-9-7-1 Query/Retrieve Setup on Vivid T8/Vivid T8 Pro
3-68
The Query/Retrieve function makes it possible to search and
retrieve DICOM data from a DICOM server for further analysis
on the Vivid T8/Vivid T8 Pro.
You may have to set up Vivid T8/Vivid T8 Pro as a destination
on the server.
1. Press F2 on the control panel and log on as administrator.
2. Select Connectivity -> Dataflow.
The Dataflow sheet is displayed.
3. Select Query retrieve from the Name pull-down menu.
Figure 3-56. Select Query Retrieve
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10