4.5.3 Selecting Test Mode

1.
In the PROGRAM screen, press the ▼ key to move the highlighted cursor to the
following position:
PROGRAM
01. TEST STEP
02. TEST MODE
03. VOLTAGE
04. HIGH LIMIT :
05. LOW LIMIT
06. ARC LIMIT
07. TEST TIME
08. RAMP TIME
09. FALL TIME
10. CHANNELS
SELECT MODE
2.
Use the [AC], [DC], [IR], [OSC] and [PA] Function Keys to select the test mode. The
test modes that are available for selection depend on the model number of the tester.

4.5.4 Description of Parameters

The parameters set in each test mode are described below.
AC Withstand Voltage (AC) Test Mode (for Models 19020/19020-4/19021/19021-4)
PROGRAM
01. TEST STEP
02. TEST MODE
03. VOLTAGE
04. HIGH LIMIT :
05. LOW LIMIT
06. ARC LIMIT
07. TEST TIME
08. RAMP TIME
09. FALL TIME
10. CHANNELS
11. SCANNER
SELECT MODE
VOLTAGE
: Sets the voltage for the AC withstand voltage test.
HIGH LIMIT
: Sets the high limit for the leakage current.
LOW LIMIT
: Sets the low limit for the leakage current. The range is less than the high
limit for leakage current or OFF.
ARC LIMIT
: Sets the ARC limit; 0 means OFF.
TEST TIME
: Sets the required test time; 0 means continuous test.
RAMP TIME
: Sets the time required for ramping to the set voltage; 0 means OFF.
FALL TIME
: Sets the time required for dropping to a low voltage; 0 means OFF.
4-16
:
1
:
AC
:
OFF
0.500
mA
:
OFF
:
OFF
:
3.0
sec OSC
:
OFF
:
OFF
:
DEFAULT
REMOTE LOCK
:
1
:
AC
:
0.050
kV DC
0.500
mA
:
OFF
:
OFF
:
3.0
sec OSC
:
OFF
:
OFF
:
DEFAULT
:
NONE
REMOTE LOCK
AC
DC
IR
PA
EXIT
CORR
ERROR
AC
IR
PA
EXIT
CORR
ERROR

Advertisement

Table of Contents
loading

This manual is also suitable for:

19020-41902119021-41902219022-4

Table of Contents