To Set the LAN
Do the following steps to set the oscilloscope to communicate with PC via LAN:
1. Connect the oscilloscope to your local area network using the network cable.
2. Press the Utility button on the front panel to enter the UTILITY function menu.
3. Press the IO Set softkey to enter the I/O SET function menu.
4. Press the Net Interface softkey, then select LAN.
5. Press the IP Set softkey to enter the IP setting interface, see the picture below:
6. Press the DHCP softkey to turn on or turn off the DHCP.
●
On: the DHCP server in the current network will assign the network parameters
(such as the IP address) for the oscilloscope.
●
Off: you can set the IP address, subnet mask, gateway manually.
➢
Turn the Universal Knob to select the desired value.
➢
Push the Universal Knob to move horizontally.
➢
Press the Save softkey to save the current settings.
7. Use NI Visa to communicate with the oscilloscope.
176 User Manual
Figure 95 LAN Settings Interface