CAN Triggering

This part will provide a brief introduction and description for the operation of the LIN
trigger.
To introduce the trigger conditions
Start— the oscilloscope will be triggered at the start bit of a frame.
Remote — the oscilloscope will be triggered by a remote frame with specified ID.
ID — the oscilloscope will be triggered by a remote or data frame that has s specified
ID.
ID+DATA— the oscilloscope will be triggered by a data frame that has a specified
ID and data.
Error — the oscilloscope will be triggered by an error frame.
Operation steps:
1. Press Setup to enter the TRIGGER function menu.
2. Press Type and select Serial.
3. Press Protocol and select CAN.
4. Press Trigger Setting to enter the CAN TRIG SET menu.
5. Press Condition and select the trigger condition using the Universal Knob:
If you select the REMOTE and ID condition:
a. Press ID Bits to set the length of the ID (11bits or 29 bits).
b. Press Curr ID Byte and use Universal Knob to select the byte that you want
to set.
c. Press the ID and set the ID's value using the Universal Knob.
(Tips: In order to make it convenient for the operator to set the parameters, ID
length is split into several bytes. For example, if the ID's length is 11 bits, it will be
split into two bytes, a byte is 8 bits. If "1st byte" is selected, only the 8 least
User Manual 105