Function Inspection
1. Press the Default button on the front panel to restore the instrument to its default
configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" under the
probe compensation signal output terminal.
T3DSO1102 scope
All scopes except T3DSO1102
3. Use the probe to connect the input terminal of CH1 of the oscilloscope to the
"Compensation Signal Output Terminal" of the oscilloscope.
4. Press the Auto Setup.
5. Observe the waveform on the display. In normal condition, the display should be a
square waveform as shown in the figure below:
6. Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform "Probe Compensation",
as detailed in the next section.
8 User Manual
Compensation Signal Output Terminal
Ground Terminal
Ground Terminal
Compensation Signal Output Terminal
Figure 7 Function Inspection