Perform Bode Plot Applicam On; Setting - Teledyne T3DSO1000A Series User Manual

Digital oscilloscope
Table of Contents

Advertisement

Perform Bode Plot Application
This section gives an overview of the steps to take when performing a Bode Plot
measurement.
1.
Press the Utility on the front panel to enter the UTILITY function menu; then press
the Next Page softkey to enter the second page.
2.
Press Bode Plot softkey to enter the BODEPLOT function menu, only the Default
button on the front panel can be used.
3.
Configure the parameters (the details see the "setting" section).
4.
Press Operate softkey to perform the test.

Setting

1.
Press Configure softkey to enter the Configure menu. All the parameters in this
menu are used to configure the external AWG.
(1)
Press AWG softkey to enter AWG menu.
Press the Amplitude softkey (Offset softkey), then turn the
Universal Knob to select the desired value, or press the Universal
Knob then enter the value using the pop up keyboard.
Press the Unit softkey, then turn the Universal Knob to select the
desired units. There are six types: Vpp, Vrms, dBV, dBu, dBm,
Arbitrary dB. The dBu is only available when the load impedance is
specified as 600 ohms.
Press the Load softkey, then turn the Universal Knob to select the
desired value, or press the Universal Knob then enter the value using
the pop up keyboard.
Figure 108 Bode plot menu
Figure 109 Bode plot configure menu
User Manual 193